Technical specifications

310A

410A

510A

620M

120CW

Tissue/Vessel size

>60 μm

>60 μm

>60 μm

>60 μm / >450 μm up to 10 mm

>60 μm

Chamber(s)

Single bath

Dual, independent or common

Dual, independent or common

Four individual chambers

Single bath/conical shape

Chamber material

Acid-resistant stainless steel

Acid-resistant stainless steel

Acid-resistant stainless steel

Acid-resistant stainless steel

Acid-resistant stainless steel

Chamber volume

Max. 10 ml

Max. 10 ml

Max. 10 ml

Max. 8 ml

Max. 10 ml

Chamber suction

No

No

No

Manual or automatic, time controlled, user defined

No

Chamber cover

Yes

With connection for suction and gassing

With connection for suction and gassing

Supplied with connections for gassing

With connection for suction and gassing

Force range

-100 -/+200 mN

-100 - /+200 mN

-100 - /+200 mN

User selectable at +/-200/400/800 mN

-100 - /+200 mN

Force resolution

0.01mN

0.01mN

0.01mN

0.01mN

0.01mN

Micropositioners

Manually operated

Manually operated

Automatical linear motor operated

Manually operated

Manually operated precision micrometer

Weight calibration

Manual

Manual

Manual

Semi Automatic

Manual

Heating

Built into chamber,
independent of superfusion
Built into chamber,
independent of superfusion
Built into chamber,
independent of superfusion
Built into chamber,
independent of superfusion
Built into chamber,
independent of superfusion

Temp. range

Ambient temp.
to 50 °C
Ambient temp.
to 50 °C
Ambient temp.
to 50 °C
Ambient temp.
to 50 °C
Ambient temp.
to 50 °C

Temp. resolution

0.1 °C

0.1 °C

0.1 °C

0.1 °C

0.1 °C

Temp. probe

External

External

External

External

External

Output reading

Force (mN)

Force (mN)

Force (mN)

Force (mN)

Force (mN)

Analog output

Up to four output, 1.0V full scale for all acquired signals, user defined
Up to four output, 1.0V full scale for all acquired signals, user defined
Up to four output, 1.0V full scale for all acquired signals, user defined

Independent filtered 4 channel output at 1.0V full scale

Up to four output, 1.0V full scale for all acquired signals, user defined

PC connection

Serial interface - RS232/RS485

Serial interface - RS232/RS485

Serial interface - RS232/RS485

USB 2.0

Serial interface - RS232/RS485

Voltage

100 to 240 VAC (auto) 50/60 Hz

100 to 240 VAC (auto) 50/60 Hz

100 to 240 VAC (auto) 50/60 Hz

External power supply

100 to 240 VAC
(auto) 50/60 Hz

Ambient temp.

15-30 °C

15-30 °C

15-30 °C

15-30 °C

15-30 °C

Optional accessories






- range

pH 0 - 14

pH 0 - 14

pH 0 - 14


pH 0 - 14

- temp. correction

0 - 50 °C

0 - 50 °C

0 - 50 °C


0 - 50 °C

 