Technical specifications |
110P/120CP |
111P |
115FP |
110PXL |
|
DMT Microscope |
No |
Yes |
No |
No |
|
Vessel size |
>60 μm |
>60 μm |
>60 μm |
>2,5 mm - 6 mm |
|
Vessel alignment |
Manually / X, Y & Z |
Manually / X, Y & Z |
Manually / X, Y |
Manually / X, Y & Z |
|
Mounting supports |
Glass pipettes |
Glass pipettes |
Glass pipettes |
Plastic tube |
|
Chamber |
Single |
Single |
Dual, one normal chamber and a composite chamber for freeze/fixative |
Single |
|
Chamber volumes |
Max. 10 ml (3 ml typical) |
Max. 10 ml (3 ml typical) |
Max. 10 ml |
Max. 10 ml (3 ml typical) |
|
Chamber material |
Acid-resistant stainless steel |
Acid-resistant stainless steel |
Acid-resistant stainless steel and POM |
Acid-resistant stainless steel |
|
Chamber cover |
With pipes for suction/gassing/superfusion |
With pipes for suction/gassing/superfusion |
With pipes for suction/gassing/superfusion |
With pipes for suction/gassing/superfusion |
|
Force range |
+/- 50 mN |
+/- 50 mN |
|
|
|
Force resolution |
0.01 mN |
0.01 mN |
|
|
|
Pressure range |
0 - 250 mmHg |
0 - 250 mmHg |
0 - 250 mmHg |
0 - 250 mmHg |
|
Pressure resolution |
0.1 mmHg |
0.1 mmHg |
0.1 mmHg |
0.1 mmHg |
|
Heating |
Built-in |
Built-in |
Built-in |
Built-in |
|
Temp.range |
Ambient - 50 °C |
Ambient - 50 °C |
Ambient - 50 °C |
Ambient - 50 °C |
|
Temp. resolution |
0.1 °C |
0.1 °C |
0.1 °C |
0.1 °C |
|
Temp. probe |
Included |
Included |
Included |
Included |
|
Weight calibration |
Semi-automatic |
Semi-automatic |
Semi-automatic |
|
|
Analog output |
0-1.0 V full scale (12 bit) |
0-1.0 V full scale (12 bit) |
0-1.0 V full scale (12 bit) |
0-1.0 V full scale (12 bit) |
|
Digital output |
Serial interface - RS232/RS485 |
Serial interface - RS232/RS485 |
Serial interface - RS232/RS485 |
Serial interface - RS232/RS485 |
|
Voltage |
100 to 240 VAC (auto) 50/60 Hz |
100 to 240 VAC (auto) 50/60 Hz |
100 to 240 VAC (auto) 50/60 Hz |
100 to 240 VAC (auto) 50/60 Hz |
|
Ambient temperature |
15-30 °C |
15-30 °C |
15-30 °C |
15-30 °C |
|
Pressure regulator |
|
|
|
|
|
Pressure range |
0 - 250 mmHg (requires external pressure source. app.1 atm.) |
0 - 250 mmHg (requires external pressure source. app.1 atm.) |
0 - 250 mmHg (requires external pressure source. app.1 atm.) |
0 - 250 mmHg (requires external pressure source. app.1 atm.) |
|
Pressure calibration |
Manual |
Manual |
Manual |
Manual |
|
Fluid capacity |
25 0ml |
250 ml |
250 ml |
250 ml |
|
Gas/superfusion ports |
Built-in |
Built-in |
Built-in |
Built-in |
|
Heating |
Built-in |
Built-in |
Built-in |
Built-in |
|
Valves |
Safety and reduction |
Safety and reduction |
Safety and reduction |
Safety and reduction |
|
Voltage |
100 to 240 VAC (auto) 50/60 Hz |
100 to 240 VAC (auto) 50/60 Hz |
100 to 240 VAC (auto) 50/60 Hz |
100 to 240 VAC (auto) 50/60 Hz |
|
Optional accessories |
|
|
|
|
|
FlowMeter |
|
|
|
|
|
- range |
15 μl/min to 1500 μl/min |
15 μl/min to 1500 μl/min |
15 μl/min to 1500 μl/min |
|
|
Enable pH meter on the interface |
|
|
|
|
|
- range |
pH 0 - 14 |
pH 0 - 14 |
pH 0 - 14 |
pH 0 - 14 |
|
- temp. correction |
0 °C - 50 °C |
0 °C - 50 °C |
0 °C - 50 °C |
0 °C - 50 °C |
|
Peristaltic pump |
2.5 - 50 rpm (for superfusion of the chamber) |
2.5 - 50 rpm (for superfusion of the chamber) |
2.5 - 50 rpm (for superfusion of the chamber) |
2.5 - 50 rpm (for superfusion of the chamber) |