Chamber |
750TOBS |
|
Chamber size(s) |
5, 10, 20 or 50 ml |
|
Chamber material |
Glass |
|
Chamber heating |
Heated air circulation (built-in - patent pending) |
|
Temp. range |
Ambient to <50 °C |
|
Temp. resolution |
0.1 °C |
|
Temp. stability |
(+/-) 0.1 °C |
|
Chamber suction |
Manual or automatic via software |
|
Chamber aeration |
Individually via precision needle valve(s) |
|
Chamber filling |
Manual/automatic via software Drain/fill/wash or constant |
|
Dimensions (w.d.h) |
55 x 50 x 50 cm |
|
Transducer |
750TOBS |
|
Force range |
(+/-) 200 mN to +/- 1600 mN (user selectable) |
|
Output resolution |
0.1 mN (using filtered signal) |
|
Transducer calibration |
Semi-automatic (via software) |
|
GLP and Maintenance |
All parts in contact with experimental agents can be replaced in <1 |
|
Reservoir |
750TOBS |
|
Reservoir volume |
4 x 900 ml or 4x1400 ml. |
|
Reservoir heating |
Electronically (built-in) |
|
Safety valve |
10 bar |
|
Temp. range |
<50 °C |
|
Temp. resolution |
0.1 °C |
|
Temp. stability |
(+/-) 0.2 °C |
|
Connectivity |
750TOBS |
|
Analog output(s) |
4 x BNC / 2.5V FS (filtered signal) or 4 x BNC / raw output |
|
Data communication |
USB (2.0) |
|
Voltage |
100 to 240 VAC (AUTO) 50/60 Hz |
|
Inlet pressure (regulated) |
Min. 2.0 bar - max 8.0 bar (internally) |
|
Air/gas |
One inlet connector on the back |
|
Vacuum |
One outlet on the back |
|
|
720MO |
|
Tissue size |
500 μm up to 10 mm as rings |
|
Chamber |
Four individual chambers |
|
Chamber material |
Acid-resistant stainless steel |
|
Chamber volume |
Max. 8 ml |
|
Chamber suction |
Manual or automatic, time controlled, user defined |
|
Chamber gassing |
Individually controlled per chamber by needle valves |
|
Chamber cover |
Supplied with connections for gassing |
|
Force range |
User selectable at +/- 200/ 400/ 800/ 1600 mN |
|
Force resolution |
0.01 mN |
|
Micropositioners |
Manually operated |
|
Weight calibration |
Semi-automatic |
|
Heating |
Built-in |
|
Temp. range |
Ambient temp. - 50 °C |
|
Temp. resolution |
0.1 °C |
|
Temp. probe |
External |
|
Output reading |
Force (mN) |
|
Analog output |
1.0 V full scale |
|
PC connection |
USB 2.0 |
|
Voltage |
External power supply |
|
Ambient temp. |
15-30 °C |
|